Chin J Schisto Control ›› 2011, Vol. 23 ›› Issue (4): 377-380.

Previous Articles     Next Articles

Diagnostic efficiency of Indirect Hemagglutination Assay Kit for antibody detection of schistosomiasis japonica

Chen Nian-gao1, Lin Dan-dan1, Xie Shu-ying1, Wang Qi-zhi2, Tang Li3, Liu Yue-min1, Zeng Xiao-jun1, Liu Hong-yun1, Huang Mei-jiao1, Chen Hong-gen1*   

  1. 1 Jiangxi Institute of Parasitic Diseases, Nanchang 330046, China|2 Anhui Provincial Institute of Schistosomiasis Control, China|3 Institute of Schistosomiasis, Hubei Center for Disease Control and Prevention, China
  • Online:2011-08-15 Published:2011-08-05
  • Contact: Chen Hong-gen

日本血吸虫病间接血凝试验诊断试剂盒临床诊断效能

陈年高1|林丹丹1|谢曙英1|汪奇志2|唐丽3|刘跃民1|曾小军1|刘红云1|黄美娇1|陈红根1*   

  1. 1 江西省寄生虫病防治研究所(南昌 330046);2 安徽省血吸虫病防治研究所;3湖北省疾病预防控制中心血吸虫病所
  • 通讯作者: 陈红根
  • 作者简介:陈年高|男|主管医师。研究方向:血吸虫病控制
  • 基金资助:

    国家自然科学基金(30960344)

Abstract:

Objective To evaluate the clinical diagnostic efficiency of Indirect Hemagglutination Assay Kit (IHA kit) for antibody detection of schistosomiasis japonica. Methods Schistosomiasis examinations by parasitological method were conducted in the schistosomiasis endemic areas in Jiangxi, Anhui and Hubei provinces, respectively. A total of 371 serum samples of schistosomiasis patients and 761serum samples of non-patients without history of schistosome infection and infested water contact were collected. All the sera were detected by using IHA Kit. The diagnostic efficiency indicators of the kit, including sensitivity, specificity, positive predicted value, negative predicted value etc. were calculated. Results The sensitivity and specificity of the kit were 95.1% (95% confidence interval was between 92.9% and 97. 3%) and 97.8% (95% confidence interval was between 96.8% and 98.8%), respectively; the positive and negative predicted values were 95.4% and 97.6%, respectively. No statistically significant difference was found in the prevalence of IHA positive detecting by the kit among the groups with different EPG levels (≤40, 41-100, >100). Also, no significant difference was observed in sensitivity and specificity among the different age groups. Conclusions The IHA kit shows a relative high efficiency for diagnosis and can be used for mass screening of schistosomiasis in field sites.

Key words: Schistosomiasis, Indirect Hemagglutination Assay kit (IHA kit), Diagnostic efficiency

摘要:

目的 评价日本血吸虫病间接血凝试验(IHA)诊断试剂盒对血吸虫病的临床诊断效能。 方法 分别在江西省、安徽省和湖北省3个疫区现场进行血吸虫病调查,并采集371份病原学确诊的血吸虫病人血清 (毛蚴孵化法或/和Kato-Katz法阳性,阳性金标准),761份无血吸虫感染史、无疫水接触史且病原学检测阴性者血清 (阴性金标准)。采用日本血吸虫病IHA诊断试剂盒对上述血清进行血吸虫抗体检测,并计算阳性符合率(灵敏度)、阴性符合率(特异度)、阳性预测值和阴性预测值等诊断效能指标。结果 日本血吸虫病IHA诊断试剂盒与阳性金标准的符合率为95.1%(95%可信区间:92.9%~97.3%), 阴性金标准的符合率为97.8%(95%可信区间:96.8%~98.8%), 阳性预测值和阴性预测值分别为95.4% 和97.6%。试剂盒对不同感染度的阳性金标准血清(1~、41~、100~)检出的抗体阳性率之间差异无统计学意义;各年龄组间IHA敏感度和特异度变化较小,差异均无统计学意义。 结论 日本血吸虫病IHA诊断试剂盒对血吸虫病具有较好的诊断效能, 可用于对血吸虫病的辅助诊断,适合应用于疫区大规模人群的血吸虫病现场筛查。

关键词: 血吸虫病, 间接血凝试验试剂盒, 诊断效能

CLC Number: